Dental Apex Locator - Dent Geosoft Endo Apex Est

$739.00 USD

GENERAL INFORMATION

  • 1.1. The device designation:
  • The device EndoEst-Apex 02 is designed for root canal working length measurement (apical constriction localization) at the time of endodontic treatment.
  • 1.2. Field of application:
  • The device is designed for use in dentistry by qualified staff. The manufacturer is not responsible for the device misuse.
  • 1.3. The device operating principle:
  • The device localizes the apical constriction by complex electrical resistance measuring between two electrodes. One of the electrodes is a metal hook (lip clip), placed on patient’s lip and the other electrode (file clip) is connected to the metallic part of the endodontic file, placed into the tooth root canal.
  • 1.4. The device advantages:
  • - Multifrequent principle of the apex localization. The measurement accuracy in any environment is guaranteed;
  • - Big and well readable LED display;
  • - Numerical, graphic, light and sound indication of measurement results;
  • - Automatic switch on when inserting in the root canal.
  • 1.5. Additional functions:
  • - The root canal humidity level indication (wet or dry);
  • - Function of apical shelf forming (“virtual apex” localization);
  • - Function of sound signals switching on/off;
  • - The device operation demo mode;
  • - Indication of power batteries discharge;
  • - Power saving function.
  • 1.6. Contraindications:
  • ! Do not use the device at those patients, who have pacemakers. 1.7. Safety measures and warnings
  • ! Use the device only with the Geosoft Dent authentic accessories (see part 3. Accessories)
  • ! Don’t dismount and don’t change anyhow the device construction. There is no element in the device control unit to be serviced by user, except changeable power batteries. In case of the device integrity damage the warranty is considered to be invalid.
  • ! In the case of long-time device storage, take off the batteries from the battery housing or shift the power switch control over to position 1 (see fig.20, page 23).
  • ! Avoid any liquid ingress into the device case.
  • ! Don’t use the device close to inflammables. The device is not operational in the presence of inflammable anesthetic mixtures with air, oxygen or nitrogen oxides
  • ! Use only sterile and disinfected device components. The device sterilization and disinfection needs to be held before the first use of the device and also after every patient in order to avoid cross infection (for more details see part 8 The device sterilization and disinfection)
  • ! At the time of working in patient’s cavity, it is strongly advised to use cofferdam.

 ACCESSORIES

  • File clip Probe Princh (3pcs / 1 pc) GЕ99.047.000 / GЕ99.121.000
  • File clip for working files. It is used as an active electrode hook (file) for apexlocation.
  • Lip clip Oral Hook (3pcs / 1pc) GЕ99.062.000 / GE99.123.000
  • It is used as passive electrode for apexlocation. It is fixed on patient’s lip.
  • Cable Signal Line ГЕ99.063.000 Cable for apexlocation. Socket—plug.
  • Accessories are in the device set and can be also shipped loose at extra cost if necessary.

 SPECIFICATIONS

  • Electrical and operational device specifications conform to the
  • European standards EN61326, EN60601-1-2.
  • Supply voltage…...………………………………….......6 (4*1,5) V
  • Electric protectability………………………………………...type В
  • Monochrome LED display……………………………...66*66 mm
  • Measurement range……………………………….from 3,5 to 0 mm
  • Numerical measurement indication……….…..from 3,5 to 0 mm
  • Indication on graphic scale……………………from 3,0 to 0 mm
  • Accuracy of measurement in the apical area……………...± 0,1 mm
  • “Virtual apex” localization……………………..from 0,1 to 0,5 mm (step 0,1 mm)
  • Time of the device operating in the mode of “inaction” until automatic transfer to “sleeping” mode/ complete device turn off……………………………………………………………10 min/5h
  • Running time of the device without batteries changing …………………………………………………..not less than 170 h
  • The control unit overall dimensions………………140*100*76 mm
  • Weight of the control unit (with power batteries) …………………………………………………not more than 365 g
  • The protection level from dust and liquid penetration…………IP51
  • The device lifetime…………………………………………..5 years